			 The LNI Schmidlin PermaCal 
			Calibration Gas Generator allows the preparation of Calibration Gas 
			Standards at a wide user-definable concentration range. 
			PermaCal uses an external source of 
			Ultra Zero Air and gas permeation tubes, combined with mass flow 
			controllers to produce user-definable gas calibration standards.  
			A gas permeation tube contains a specific substance in liquid and 
			vapour phases.  The diffusion rate of the vapour phase through 
			the tube is a function of temperature.  By maintaining the tube 
			at a constant controlled temperature, the gas permeation rate 
			becomes constant.  Using mass flow controllers to generate a 
			range of dilution flows of Ultra Zero Air, PermaCal can generate 
			specific concentrations of calibration gas. 
			Applications: 
			
			Features: 
			
				Permeation oven: accepts up to 6 permeation tubes (up to 12 cm long)  
				Oven Temperature: 40 - 80 oC  
				Permeation oven flow: 100 mL/min  
				Total flow: 500 to 5,000 mL/min (depending on concentration setting)  
				Requires external source of Ultra Zero Air  
				Inlet pressure: 45 psig  
				Outlet pressure: atmospheric pressure  
				Small size (6" W x 14" D x 15" H) conserves lab bench space  
				LCD touch-screen display provides simplified user interface  
				Weight: 31 lbs  
				1 year parts & labour warranty  